Characterisation and mapping of scattered radiation fields in interventional radiology theatres (2024)

Introduction

Proper evaluation of worker exposure is a major issue and current challenge in radiation protection. The effectiveness of worker protection greatly depends on an understanding of the radiation field to which the staff is exposed. A radiation field can be described by fundamental physical quantities: the particle fluence and the energy spectrum. Besides physics, evaluating worker exposure is based on current knowledge of the biological effects of ionizing radiation. The concept of equivalent dose for any given organ and effective dose was introduced for risk management purposes as well as to establish annual dose limits1. The effective dose cannot be measured. The calculation of the effective dose integrates factors related to biological effects observed in radiobiological and epidemiological studies2.

The continuous improvement in understanding the underlying mechanisms of the health effects due to exposure to ionizing radiation motivates a regular re-evaluation of the annual limits.

In 2011, the International Commission on Radiological Radioprotection (ICRP) further restricted the annual eye lens dose limit, resulting in changes to the legal limits in Europe and Switzerland2. Before 2011, the annual dose limit for the eye lens was 150mSv, this limit was set in order to prevent possible cataracts at the end of a career. It was agreed by consensus that if the annual effective dose limit was respected for the whole body it would also imply compliance to the eye lens dose limit3. The eye lens dose limit being relatively high, the probability of exceeding it was then quite low. Following a re-evaluation of the effects on the Hiroshima and Nagasaki cohort4, the limit was then lowered from 150 to 20mSv5. The division of the limit by a factor of 7 significantly changes the stakes, as a large number of staff members may be over this limit. Because of the introduction of this more restrictive limit, a new dedicated operational quantity (Hp(3)) was then proposed for managing eye lens exposure6. This prompted strong debate on the methods and means of measurements as well as calibration procedures for the measuring instruments.

The 2017 ICRP report also points out the need to revise the operational quantities for external exposure7. The ICRU sphere is not used anymore; direct conversion coefficients are instead given to the operational quantities from the fluence using anthropomorphic reference computational phantoms. In line with this new approach, we decided to characterise the radiation field in an interventional radiology (IR) theatre.

Our second reason for wanting to characterise the radiation field is related to the evolution of imaging systems. Medical imaging experts are permanently optimising the dose received by patients8, as well as using increasingly sophisticated imaging systems and improved image reconstruction algorithms. Additional filters are used to optimise the procedures (patient morphology, paediatric, etc.). These changes influence the energy spectrum used to image the patient’s body and thus the energy spectrum of the scattered radiation to which the medical staff is exposed. These changes are often accompanied by a decrease of the dose received by the hospital staff per procedure. Recent studies show that the development of more complex and longer procedures combined with the higher number of patients inevitably leads to an increase in exposure of hospital staff members9. Most procedures require staff to remain close to the patient during the image-guided procedures. A study by Sanchez et al.10 carried out on radiologists shows that more than 50% of them receive a dose higher than the authorized annual limit of 20mSv, sometimes going up to monthly doses higher than 20mSv on apron. This means it is important to establish an accurate dosimetry surveillance program in order to avoid underestimating worker dose. Knowledge of the spectrum to which the medical staff is exposed would enable a more accurate calculation of the dose, but would also make it possible to retrospectively know the dose received by an organ and this despite the evolution over time of the operational quantities for external exposure.

The spectral composition of X-ray radiation has been studied for a long time. Although many studies have been carried out based on computer simulations of the spectrum, very few are based on direct measurements. The first studies of the spectral composition of beams from an X-Ray tube date back to the mid-30’s11,12. The X-ray spectral information was first investigated by measuring attenuation curves. Forty years later, the first computer simulations on primary radiation were performed13,14,15. The first direct measurements of the spectrum were in 1969 by Peaple and Burt16, later followed by Seelentag and Panzer17 and Birch and Marshall18 using crystal scintillator detectors (NaI) or a semiconductor detector (Ge). The measurement set-up leads to cumbersome experimental conditions and involved strong constraints such as concomitant cooling or reducing the counting rate with a distance of several meters to the source and high collimation. In 1983, Castro et al.19 attempted to solve those problems, using a CdTe detector with a low detection efficiency. However, this kind of measurement requires long-term work along with a large number of corrections and is not suitable for rapid cumulative dose measurements for a procedure or a configuration. More recently, Miyajima et al.20 used a CdZnTe detector for spectra measurement, with results compatible with those obtained by Monte-Carlo (MC) simulations. The detector has several problems such as poor charge transport properties and requires complex calculated corrections for the spectral distortion.

Adapting a detector to a routine dosimetry application would imply that the device could be used regardless of the radiation field, the collimation, or the position in the radiation field. A very challenging point is the acquisition speed of the data. Missing data can lead to a wrong estimation of the photon fluence and dose. Another challenge to negotiate is making the detector small enough to be comfortable for the user to wear.

With the introduction of new technologies, the measurement of the energy spectrum in hospital theatres can provide optimised radiation protection safety for workers. Since 1997, the CERN Medipix team has been developing novel photon detectors that are able to provide spatial, temporal, and energy information for each incoming photon. This means that these detectors are able to measure the energy spectrum in real time. In this work, we used the Timepix3 detector to measure the energy spectrum of the scattered radiation field in an IR theatre. Compared to other spectral detector16,17,18,19,20, this detector has the advantage of coping with higher photon fluences and not requiring beam collimation with heavy pinhole shielding. It can be used at any distance from the radiation source in the scattered field and data are available instantaneously. This enabled us to compare the different spectra for different positions in the room, and for different heights. We then created a map of the scattered radiation field extracted directly from the energy spectra as received by a staff member around the X-ray tube.

Materials and methods

Timepix3 detector and software

The Timepix3 chip is a hybrid pixel detector (HPD) developed in the framework of the Medipix3 collaboration at CERN21, and is a latest version of the Timepix family22. We used the Timepix3 chip with a 500μm thick silicon sensor produced and bump bonded by ADVACAM (Prague, Czech Republic)23. We choose a Si sensor layer despite a low absorption efficiency (4% detection at 60keV versus 87% for a 500µm CdTe sensor)24 as this is counterbalanced by a more stable Si response since the production of fluorescence photons is much higher in CdTe, from 80 to 90% versus less than 5% for Si. With a mean free path of fluorescence photons of 10μm with Si and 110μm with CdTe. The chip has a pixel matrix of 256 × 256 square pixels with 55μm side and can process up to 40 Mhits/cm2/s. Each pixel sensor is linked to his own readout. A 500nm aluminium layer covers the sensor. The sensor was biased at 300V. The data driven mode was selected to collect information on arrival time, charge deposit, and the coordinates of each pixel hit by a photon. The charge deposit was calculated based on the time over threshold value. Each pixel was calibrated independently using radioactive sources and fluorescence peaks of different material foils25. We choose to analyze data with signal collected with single pixels only. The choice was made to eliminate additional uncertainties on energy calculation for photons generating a signal on several pixels (pixel sharing correction). Single pixels data do indeed represent only a fraction of all detected photons. The data have thus been normalized taking into account the ratio between the number of photons with single pixel and multiple pixel signals. The overall correction includes the detector efficiency as well as for angular variation, temperature, and light variation, since calibration condition can be different than measurement conditions26. Uncertainties have been calculated for energy calibration, pixel sharing, detector efficiency, position, angular variation, temperature, and light variation.

In this study, the Timepix3 chip was connected to the AdvadDAQ (Advacam, Prague, Czech Republic) readout system and data was collected using the Pixet software, version 1.5.0.714 (Advacam, Prague, Czech Republic).

Measurement in interventional radiology theatre

All measurements were carried out in an IR theatre at Christchurch Hospital—New Zealand, using an X-ray imaging system ALLURA XPER FD 20 (Philips, Hamburg, Germany). An anthropomorphic male phantom (RSD, Long Beach, CA, USA) was used to simulate a patient lying on the radiotransparent carbon fiber table. Energy spectra of the scattered radiation field were acquired around the table with the Timepix3 detector. Measurements were taken in a 3m radius half-sphere centered on the abdomen of the phantom and at four different heights corresponding to the eye lens, chest, belt and knee, respectively 170cm, 135cm, 96cm and 53cm of a 1.76m medical staff member (Fig.1). The chosen protocol was the default set-up for pelvis/iliac examinations with an image acquisition rate of 3 frames per seconds (fps), a tube voltage of 74kVp with a current of 12.0mA, using the mode of standard dose for a reference patient (70kg), leading to a filtration of 0.4mm Cu + 1mm Al. The cumulative time of measurements depended on the position in the room and ranged from 1min for the closest positions to 4min for the furthest away. Each data acquisition point was repeated 3 times; in this article we will present only the mean value of the three measurements for each position. All data have been normalized on 4min measurement.

Measurements taken at four different heights in the IR theatre.

Full size image

A first series of measurements consisted in measuring the radiation field at the standard positions where the staff members stand during a routine procedure. After a period of observation and discussion with the team in place, we identified the four most common positions (see Fig.2). Collective protection gear (suspended lead screen and lead table drape) were systematically used. We held the Timepix3 chip by hand above the lead apron, and we performed the measurements at the four heights successively in the four positions where the staff stand during the procedures, identified as A, B, C and D in the theatre on Fig.2.

Left: positions of the different staff members in the IR room during procedure. Right: diagram of the different measurement positions (represented by red or blue stars) to create the scattered radiation map of the IR theatre.

Full size image

A second series of measurements consisted in measuring the energy spectra around the patient couch. Acquisitions were taken every 15°, on 180°, around the anthropomorphic phantom, every 50cm (Fig.2, right), for the four heights mentioned previously. The cumulative time of measurements depended on the position in the room and ranged from 1min for the closest positions to 4min for the furthest away. All data have been normalized on 4min measurement. The Timepix3 chip was set on a metallic stand infusion holder. No specific means of protection were used. Moreover, the effect of the radiation protection gear (lead screen and lead table drape) was tested at a few positions in the room. On Fig.2 right, each intersection between an arc and a line represents a position of measurement. The configuration of the room (walls, furniture) did not allow us to do all positions. Measurement points are represented by stars (blue or red). All measurements were performed after the surgery planned for the day. No measurement was done with a patient present in the theatre. These measurements were used to create a map of the theatre. Data has been corrected for pixel sharing and detection efficiency. Each spectrum obtained was segmented into energy bins, each multiplied with an energy-specific weighting factor. The values represented on the map are interpolated with a linear function between the measurement points.

Ethical approval

We confirm that all the methods in this study have been applied in accordance with national guidelines and regulations. The consent of each participant was obtained prior to measurements. Each participant was informed and aware of the risks associated with ionizing radiation. Experimental protocols were approved by CERN Radiation Protection group.

A written consent to publish was obtained by all human participants involved in the study.

Results

Figures3 and 4 show the energy spectrum measured on a person holding the Timepix3 chip at four heights in the four standard positions of the medical staff. Both figures present the same dataset, in a different way. Figure3 presents the spectrum measured for a given height for each staff member. On Fig.3 scales have been adapted to make the comparison easier. Figure4 shows the energy spectrum measured at four heights for a given staff member. Scales are position dependent.

Energy spectra for each medical staff at a given height, with collective protections. Light curves represent the uncertainty.

Full size image

Energy spectra in absolute values for each height for a given person, with collective protections. Light curves represent the uncertainty.

Full size image

Figure5 shows the different energy spectra obtained with and without protective gear. Figure6 shows the ionizing radiation levels at the 4 different heights in the IR theatre. Figure7 represents the fluence as a function of the height, 50cm away from the X-ray tube for an angle of 105° without protection gear. The energy spectrum is given for each of the four heights, i.e. eye lens, chest, belt, and knee.

Energy spectra at 50cm of the X-ray tube for an angle of 105° at 135cm of height. Blue line represents the spectra without radiation protection gear, the orange line represents the spectra obtained with a lead screen and a lead table shield, the yellow line represents the data of the orange spectrum multiplied by 130. The dotted line represents the average energy.

Full size image

On the left part, the fluence multiplied by a factor energy-dependent as a function of the height, on the right part the energy spectra corresponding for each different height (from top to bottom: 170cm, 135cm, 96cm, 53cm). Eav represent the average energy.

Full size image

Discussion

The available data on direct measurements of the scattered radiation field in IR is sparse. An overview of the literature indicates that almost all the studies are done with Monte Carlo simulations13,14,15. Our measurements provide information on the radiation fluence in the theatre as well as the energy spectrum of the scattered photon field. Those two combined pieces of information are helping to understand the exposure level of the medical staff and to evaluate the homogeneity of the field to which the different people are exposed. This approach is part of the needs of the ICRP and ICRU for fluence-based dosimetry7. The spectra presented here clearly show that we should need consideration of energy dependence of dosimeter to evaluate doses (Figs.3 and 4). Since the X-ray spectrum changes depending on the measurement positions (Figs.3 and 4), it is important to pay attention to the estimation procedure when attempting to estimate the exposure dose at other locations using the value of the exposure dose measured at a given position. Studying the position of the dosimeter is crucial to avoid underestimation of the dose measurement.

The energy threshold for the Timepix3 is 3keV due to the calibration curve. Marshall et al.27 measured the energy spectrum with a germanium detector with a higher detection threshold set at 20keV, and Kalyvas et al. obtained a higher threshold set at 10keV with an Amptek XR-10 CdTe28.

The person in position A is the most exposed to ionizing radiations because of his or her proximity to the source of scattered radiation (i.e. the entry point of the primary beam in the patient body), even if standing right behind the protective shields (Fig.3). The least exposed person is the one standing in the back of the theatre, at position D in Fig.3.

In Fig.4, the person at position A is less exposed at the level of the eye lens compared to the rest of the body. This can be explained by the fact that the person stands right behind the ceiling protective panels. Collected data also helps to highlight a shift in energy from head to toe at position A, indicating a non-homogenous exposure of the radiologist. The lower exposure level is observed at the level of the eye lens and the shift in energy indicates that if the eye lens dosimetry is performed using a dosimeter worn on the chest, the dose might be overestimated. A dedicated dosimeter would thus be recommended for a better monitoring.

Energy spectra show a more homogeneous exposure for staff standing in the back of the theatre, the spectra shape and fluence are very similar for eye lens, chest and belt. This can be explained by the fact that people stand farther away from the patient who generates the scattered radiation field. For those people, a dosimeter worn on the chest would provide a measurement that is representative of the whole body and of the eye lens exposure. Measured fluence above 74keV can be due to pile up effect.

The highest exposure for position A and D is observed at the level of the knee. The same pattern was observed by Rehn29 with a higher dose close at the knee level when using a C-arm tube, with the tube under the patient couch. An area with a higher scatter exposure rate is found close to the patient, in accordance with Fig.6. The exposure of staff across a theatre shows that a dosimeter worn at chest level can overestimate the exposure of other organs (e.g. position C) or underestimate it (position A and B). The majority of workers are far from the annual dose limits. On the other hand, this could be a problem for a small fraction of workers, notably radiologists and cardiologists, who may reach dose limits on a dosimeter worn on the chest. Moreover, the articles of Ciraj-Bjelac et al.30 and Hupe et al.31 show that performance of many detectors decrease by using in pulsed field.

For pedagogical purposes, we investigated the effectiveness of the radiation protection gear. The use of a lead screen as well as a lead table shield reduces by more than a factor 130 the ionizing radiation field at chest level of a person standing 50cm away from the tube at an angle of 105°. A spectrum hardening due to absorption by the protective material is clearly observed with a shift from a mean energy of 44keV without the gear to 50keV with gear (Fig.5). Mori and McCaffrey32,33 show that this difference in the mean energy can lead to a 12% higher X-ray transmission rate of protective aprons. Figure5 was of particular interest when we presented the results to the staff and raised questions. It helped to increase awareness and understanding about radiation protection principles. In fact, the medical staff were interested to know how they could adapt their personal radiation protection gear, their positions in the room, or how to move or adapt the collective radiation protection gear. Medical staff were able to see and quantify the effectiveness of the protection, or the most irradiating position in the room. The radiation protection training program for the staff in IR in Lausanne University Hospital uses the DoseAware tool, and it is also very much appreciated by all participants. The Timepix chip shows great potential for offering further options and bringing additional information compared to other pedagogical tools available on the market. The Timepix3 chip provides information on the energy spectrum and opens possibilities to strengthen understanding of effective radiation protection. However, spectra presented are those measured by the sensitive layer of the detector and are not the energy spectra of the scattered radiation field in air. Full deconvolution of the spectrum will be needed before to relate to a dose. Experiences in classrooms have already demonstrated that the Timepix chip could be a valuable educational tool for learning about radiation34,35. Our experience demonstrates that the use of the chip as a pedagogical tool can be extended to the medical environment.

All data collected in the theatre was used to generate a coloured distribution map depicting the exposure levels in horizontal planes at four heights. As expected, the highest exposure levels were, for all four planes, next to the X-ray tube. The maximum exposure level was recorded on the horizontal plane at 135cm (at chest level) on the left side of the tube, i.e. the side that is not covered by the protective gear. The radiologist stands on the right side, behind the gear. The exposure is higher on the two planes under the patient’s bed where the tube is placed, as illustrated on the training video of the Federal office of public health in Switzerland36. The radiation distribution represents a specific distribution for the theatre in which we did the measurements and is an innovative and useful tool for the local radiation protection officer. The limitation of this study is the lack of measurements for other angles of the X-ray system.

Figure7 shows that exposure to the body is height-dependent, therefore varies for the organs. The highest exposure for a person standing on the red star (see Fig.2, right) is found at chest level. The shapes of the energy spectra are also different with the height. The maximum energy of the scattered radiation is higher at higher heights. Also, if the fluence for the eye lens is lower than for the rest of the body, the mean energy is higher. This information can be important since currently a chest dosimeter is considered sufficient to estimate eye lens exposure. But from this study we noticed the difference between all the energy spectra for different body parts. This confirms that the choice of the dosimeter position is important for an efficient dosimetry surveillance program. A dosimeter worn on the chest may not be representative of the eye lens exposure because of a different radiation exposure level and different energy spectrum.

If we consider the current dosimeters frequently used: TLD and OSL, their response in energy is not constant. TLDs present an energy response ranging from 0.8 to 1.5 up to 1000 keV37, while OSL has a linear decreasing response (from 1.07 to 0.85) between 0 and 120 keV38. The energy chosen for the calibration of the detector is important, as it is often very different from the energy received monthly. A correction factor must be taken into account39. However, this factor is based on an average theoretical energy received and not on the real exposure. In this sense, the Timepix3 brings a new performance and makes it possible to directly calibrate the detector for every possible energy.

Also, the energy correction factor may be different for the chest and the eye lens. A dedicated dosimeter may thus be recommended for eye lens exposure monitoring. Moreover, the radiation field becomes more homogenous the further from the patient while the dose rate decreases. The information collected with the Timepix3 chip in the theatre may also be important with respect to the new approach for dosimetry based on direct conversion coefficients given to the operational quantities from the fluence and energy. This provides us with new parameters for the radiation protection of medical staff.

Conclusion

Here we have demonstrated the feasibility of using the energy-resolving hybrid pixel detector, Timepix3, to measure X-ray information and characterize the scattered radiation field in an IR theatre. The radiation fields were characterised by measuring the fluence and energy spectra at different positions in an IR theatre. This new approach makes it possible to compare the different energy spectra to which staff members are exposed and thus guide the radiation protection officer in choosing appropriate radiation protection gear. As expected, the radiation fields to which members are exposed vary significantly with their position in the room. Moreover, for a given person, this new radiation detection technique provides information regarding the part of the body that is most exposed. The detector could also be used to correct practitioner practices, for example by accurately positioning the lead screens to improve their protection. Finally, the Timepix3 has demonstrated a great potential as a pedagogical tool to enhance understanding of the ionizing radiation hazards in an interventional theatre.

References

  1. IXRPC. International recommendations for X-ray and radium protection. Br. J. Radiol. 7, 695–699 (1934).

    Article Google Scholar

  2. Stewart, F. A. et al. ICRP publication 118: ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs–threshold doses for tissue reactions in a radiation protection context. Ann. ICRP 41, 1–322 (2012).

    Article CAS Google Scholar

  3. Ciraj-Bjelac, O. et al. Occupational exposure of the eye lens in interventional procedures: how to assess and manage radiation dose. J. Am. Coll. Radiol. 13, 1347–1353 (2016).

    Article Google Scholar

  4. Minamoto, A. et al. Cataract in atomic bomb survivors. Int. J. Radiat. Biol. 80, 339–345 (2004).

    Article CAS Google Scholar

  5. ICRP 2011 Statement on Tissue Reactions (n.d.)

  6. Behrens, R. On the operational quantity Hp(3) for eye lens dosimetry. J. Radiol. Prot. 32, 455–464 (2012).

    Article CAS Google Scholar

  7. ICRP. Draft Joint Report of ICRU and ICRP for consultation: operational quantities for external radiation exposure.

  8. Racine, D. et al. Objective comparison of high-contrast spatial resolution and low-contrast detectability for various clinical protocols on multiple CT scanners. Med. Phys. 44, e153–e163 (2017).

    Article Google Scholar

  9. FOPH. Audits in the OR-areas of Swiss Hospitals. (2020).

  10. Sánchez, R. M. et al. Staff doses in interventional radiology: a national survey. J. Vasc. Interv. Radiol. 23, 1496–1501 (2012).

    Article Google Scholar

  11. Silberstein, L. XXXIV. Spectral composition of an X-ray radiation determined from its filtration curve. Lond. Edinb. Dublin Philos. Mag. J. Sci. 15, 375–394 (1933).

    Article CAS Google Scholar

  12. Mayneord, W. V. & Roberts, J. E. The “quality” of high voltage radiations. I.. Br. J. Radiol. 8, 341–364 (1935).

    Article Google Scholar

  13. Twidell, J. W. The determination of X-ray spectra using attenuation measurements and a computer program. Phys. Med. Biol. 15, 529 (1970).

    Article CAS Google Scholar

  14. Sundararaman, V., Prasad, M. A. & Vora, R. B. Computed spectra from diagnostic and therapeutic X-ray tubes. Phys. Med. Biol. 18, 208 (1973).

    Article CAS Google Scholar

  15. Stanton, L., Lightfoot, D. A. & Mann, S. A penetrameter method for field kV calibration of diagnostic X-ray machines. Radiology 87, 87–98 (1966).

    Article CAS Google Scholar

  16. Peaple, L. H. J. & Burt, A. K. The measurement of spectra from X-ray machines. Phys. Med. Biol. 14, 73 (1969).

    Article CAS Google Scholar

  17. Seelentag, W. W., Panzer, W., Drexler, G., Platz, L. & Santner, F. A Catalogue of Spectra for the Calibration of Dosemeters. https://inis.iaea.org/Search/search.aspx?orig_q=RN:11531730. (1979).

  18. Birch, R. & Marshall, M. Computation of bremsstrahlung X-ray spectra and comparison with spectra measured with a Ge(Li) detector. Phys. Med. Biol. 24, 505–517 (1979).

    Article CAS Google Scholar

  19. Castro, E. D., Pani, R., Pellegrini, R. & Bacci, C. The use of cadmium telluride detectors for the qualitative analysis of diagnostic X-ray spectra. Phys. Med. Biol. 29, 1117–1131 (1984).

    Article Google Scholar

  20. Miyajima, S., Imagawa, K. & Matsumoto, M. CdZnTe detector in diagnostic x-ray spectroscopy. Med. Phys. 29, 1421–1429 (2002).

    Article CAS Google Scholar

  21. Medipix collaboration. Medipix website. Medipix (2017).

  22. Llopart, X., Ballabriga, R., Campbell, M., Tlustos, L. & Wong, W. Timepix, a 65k programmable pixel readout chip for arrival time, energy and/or photon counting measurements. Nucl. Instrum. Methods Phys. Res. 581, 485–494 (2007).

    Article ADS CAS Google Scholar

  23. ADVACAM. ADVACAM: Imaging the unseen. https://advacam.com/.

  24. Ballabriga, R. The Design and Implementation in 0.13um CMOS of an Algorithm Permitting Spectroscopic Imaging with High Spatial Resolution for Hybrid Pixel Detectors. (Universitat Ramon Llull, 2009).

  25. Jakubek, J. Precise energy calibration of pixel detector working in time-over-threshold mode. Nucl. Instrum. Methods Phys. Res. 633, S262–S266 (2011).

    Article ADS CAS Google Scholar

  26. Nowak, M., Tlustos, L., Carbonez, P., Verdun, F. R. & Damet, J. Characterisation of the impacts of the environmental variables on Timepix3 Si sensor hybrid pixel detector performance. Nucl. Instrum. Methods Phys. Res. 981, 164502 (2020).

    Article CAS Google Scholar

  27. Marshall, N. W., Faulkner, K. & Warren, H. Measured scattered x-ray energy spectra for simulated irradiation geometries in diagnostic radiology. Med. Phys. 23, 1271–1276 (1996).

    Article CAS Google Scholar

  28. Kalyvas, N. Measurement of scatter radiation spectrum from radiographic units. Med. Phys. https://doi.org/10.1594/ECR2013/C-0309 (2013).

    Article Google Scholar

  29. Rehn, E. Modeling of Scatter Radiation During Interventional X-ray Procedures (Linköping University, Linköping, 2015).

    Google Scholar

  30. Ciraj-Bjelac, O., Carinou, E. & Vanhavere, F. Use of active personal dosimeters in hospitals: EURADOS survey. J. Radiol. Prot. 38, 702–715 (2018).

    Article CAS Google Scholar

  31. Hupe, O., Friedrich, S., Vanhavere, F. & Brodecki, M. Determining the dose rate dependence of different active personal dosemeters in standardized pulsed and continuous radiation fields. Radiat. Prot. Dosimetry https://doi.org/10.1093/rpd/ncz173 (2019).

    Article PubMed Google Scholar

  32. Mori, H., Koshida, K., Ishigamori, O. & Matsubara, K. Evaluation of the effectiveness of X-ray protective aprons in experimental and practical fields. Radiol. Phys. Technol. 7, 158–166 (2014).

    Article Google Scholar

  33. McCaffrey, J. P., Tessier, F. & Shen, H. Radiation shielding materials and radiation scatter effects for interventional radiology (IR) physicians: Radiation shielding materials and radiation scatter for IR. Med. Phys. 39, 4537–4546 (2012).

    Article CAS Google Scholar

  34. Bott, O. J., Dresing, K., Wagner, M., Raab, B.-W. & Teistler, M. Informatics in radiology: use of a C-arm fluoroscopy simulator to support training in intraoperative radiography. RadioGraphics 31, E65–E75 (2011).

    Article Google Scholar

  35. Loy Rodas, N. A global radiation awareness system using augmented reality and Monte Carlo simulations. RadioGraphics https://doi.org/10.1594/ECR2018/C-1615 (2018).

    Article Google Scholar

  36. O. fédéral de la santé publique. Vidéos de formation pour les établissements médicaux. https://www.bag.admin.ch/bag/fr/home/gesetze-und-bewilligungen/gesuche-bewilligungen/bewilligungen-aufsicht-im-strahlenschutz/informationen-fuer-medizinische-betriebe/schulungsfilme-medizinische-betriebe.html.

  37. Duggan, L., Hood, C., Warren-Forward, H., Haque, M. & Kron, T. Variations in dose response with x-ray energy of LiF:Mg, Cu, P thermoluminescence dosimeters: implications for clinical dosimetry. Phys. Med. Biol. 49, 3831–3845 (2004).

    Article CAS Google Scholar

  38. Al-Senan, R. M. & Hatab, M. R. Characteristics of an OSLD in the diagnostic energy range: characteristics of OSLDs in diagnostic energy range. Med. Phys. 38, 4396–4405 (2011).

    Article Google Scholar

  39. Asahara, T. et al. Exposure dose measurement during diagnostic pediatric X-ray examination using an optically stimulated luminescence (OSL) dosimeter based on precise dose calibration taking into consideration variation of X-ray spectra. Radiat. Meas. 119, 209–219 (2018).

    Article CAS Google Scholar

Download references

Acknowledgement

The authors would like to thank the Medipix3 collaboration. Authors are grateful to the interventional radiology team of Christchurch Hospital—NZ, for their kindness, technical support, and help during measurements.

Author information

Authors and Affiliations

  1. CERN, European Organization for Nuclear Research, Geneva, Switzerland

    M. Nowak,P. Carbonez&J. Damet

  2. Institut of Radiation Physics, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland

    M. Nowak,F. R. Verdun&J. Damet

  3. Department of Radiology, University of Otago, Christchurch, New Zealand

    P. Carbonez,M. Krauss&J. Damet

  4. Department of Interventional Radiology, Christchurch Hospital, Christchurch, New Zealand

    M. Krauss

Authors

  1. M. Nowak

    View author publications

    You can also search for this author in PubMedGoogle Scholar

  2. P. Carbonez

    View author publications

    You can also search for this author in PubMedGoogle Scholar

  3. M. Krauss

    View author publications

    You can also search for this author in PubMedGoogle Scholar

  4. F. R. Verdun

    View author publications

    You can also search for this author in PubMedGoogle Scholar

  5. J. Damet

    View author publications

    You can also search for this author in PubMedGoogle Scholar

Contributions

M.N. wrote the main manuscript text. M.N., P.C., F.R.V., J.D. conceived and designed the analysis. M.N., P.C., J.D. collected the data. M.K. contributed data and analysis tools. M.N. performed the analysis. All authors reviewed the manuscript.

Corresponding author

Correspondence to M. Nowak.

Ethics declarations

Competing interests

The authors declare no competing interests.

Additional information

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Characterisation and mapping of scattered radiation fields in interventional radiology theatres (8)

Cite this article

Nowak, M., Carbonez, P., Krauss, M. et al. Characterisation and mapping of scattered radiation fields in interventional radiology theatres. Sci Rep 10, 18754 (2020). https://doi.org/10.1038/s41598-020-75257-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1038/s41598-020-75257-5

Characterisation and mapping of scattered radiation fields in interventional radiology theatres (2024)
Top Articles
Latest Posts
Recommended Articles
Article information

Author: Rob Wisoky

Last Updated:

Views: 5975

Rating: 4.8 / 5 (48 voted)

Reviews: 87% of readers found this page helpful

Author information

Name: Rob Wisoky

Birthday: 1994-09-30

Address: 5789 Michel Vista, West Domenic, OR 80464-9452

Phone: +97313824072371

Job: Education Orchestrator

Hobby: Lockpicking, Crocheting, Baton twirling, Video gaming, Jogging, Whittling, Model building

Introduction: My name is Rob Wisoky, I am a smiling, helpful, encouraging, zealous, energetic, faithful, fantastic person who loves writing and wants to share my knowledge and understanding with you.